Radiometer

Summary



TypeRadiometer
VendorThe Eppley Laboratory, Inc.
ModelEppley Precision Infrared Radiometer Pyrgeometer
S/NPIR35116F3; PSP35149F3
ID Prefixpyrgeo
CORIOLIX IDepprmr_rad_000101
ClassAtmospheric
LocationCrane Cab
Conditionunknown
Configurationdefault
Native Sample Rate0.25 Hz
UDP Port(s)17012 16000
EnabledTrue
 
 
 

General Description



Two radiometers measuring irradiance include a pygeometer (PIR) measuring downwelling longwave irradiance and a pyranometer (PSP) measuring global shortwave irradiance.

 
 

Data Collection Specs

Native Data Feed
Signal Type
Communication Type
Serial Details
Baud Rate 19200
Parity N
Stop Bits 1
Byte Size 8
Timeout (s) None
Polled Start Commands
Polled Serial Commands
Ethernet Details
MAC Address
IP Address
Protocol
Source File Details
Path
Filename
Native Data Format
Native Data Description
Native Data Type text
Start of Message Characters
End of Message Characters
Sample Data Message $WIR01,19/01/17,19:18:32, 67, -285.2, 216.54, -0.44, -0.05, 405.76, 1.4, 10.9
Message Parser Format ['^\WWIR01,(?P<instrument_date>\d\d/\d\d/\d\d),(?P<instrument_time>\d\d:\d\d:\d\d),(?P<test>[\w])?\s*(?P<samples_in_average>\d+),\s+(?P<pir_volts>\-?\d+\.\d+),\s+(?P<pir>\-?\d+\.\d+),\s+(?P<pir_case_temp_false>\-?\d+\.\d+),\s+(?P<dome_temp>\-?\d+\.\d+),\s+(?P<psp>\-?\d+\.\d+),\s+(?P<diagnostic_rad_board_temp>\-?\d+\.\d+),\s+(?P<diagnostic_rad_board_vbat_voltage>\-?\d+\.\d+)$']
Data Processing
Data Processing Script NoTransformation
Data Rates
Native Sample Rate 0.25 Hz
Shipboard Network Packet Rate 0.25 Hz
Ship to Shore Rate 0.25 Hz
Data Storage & Access
Raw Data Ethernet Distribution
IP Address
Port 17012
Packet Size None
Transformed Data Ethernet Distribution
IP Address 192.167.0.255
Port 16000
Archive to Database True
Transmit Real-time Data to Shore True
Time Source
Source
 

Parameters


Quantity Infrared Irradiance (W/m2) Shortwave Irradiance (W/m2)
Plot Color
 
 
Parameter Full Name Infrared Irradiance Shortwave Irradiance
Parameter Short Name Infrared Irradiance Shortwave Irradiance
Parameter Standard Name XXXXXXXX XXXXXXXX
SAMOS Data Category Longwave Radiation Shortwave Radiation
Description Pyrgeometer measures the downwelling longwave (infrared) irradiance from the sky. The spectral range covered is 4 to 50 µm. The accuracy is 5 W/m2. Pyranometer (PSP) mounted in the horizontal plane to measure shortwave irradiance from 285 to 2800 nm.
Processing Status raw raw
Processing Symbol pir psp
Diagnostic False False
IOOS Category Meteorology Meteorology
Units W/m2 W/m2
Data Type double double
Gross Minimum Value None W/m2 None W/m2
Gross Maximum Value None W/m2 None W/m2
Recommended Minimum Value None W/m2 None W/m2
Recommended Maximum Value None W/m2 None W/m2
Global Minimum Value 0.0 W/m2 0.0 W/m2
Global Maximum Value 500.0 W/m2 1400.0 W/m2
Local Minimum Value None W/m2 None W/m2
Local Maximum Value None W/m2 None W/m2
Lower Detection Limit 2.0 W/m2 2.0 W/m2
Saturated Value 500.0 W/m2 1400.0 W/m2
Bad Data Value 500.0 W/m2 1400.0 W/m2
Resolution None W/m2 None W/m2
Accuracy +/- None W/m2 +/- None W/m2
Precision None W/m2 None W/m2
Binning Type linear linear
Online Display Format 5.1f 6.1f
Logging Modes
 

Physical Sensor Specs

Operating Limits
Input Power None - None V AC
Ideal Power None V AC
Operating Current None - None mA @ None V AC
Analog Output None - None
Operating Flow Rate None - None ml s-1
Ideal Flow Rate None ml s-1
Operating Depth None - None m
Operating Temperature None - None degree_C
Lamp Life None days
Response Time None seconds
Additional Time Offset 0.0 seconds
Physical Status
Current Location installed
Current Status in use
Physical Condition unknown
Physical Configuration default
Software Version
Software Configuration
Data Acquisition Device ID
Data Acquisition Device Serial Port
Data Acquisition Device Analog/Digital Port(s)
Data Acquisition Device Modbus Unit None
Sensor Use Level
Enabled True
Vessel Installation Location
Vessel Endeavor
Vessel Location Crane Cab
Location Coordinates (meters)
Sensor Orientation vertical up
Height Above Waterline (meters) None
Zero Line Reference (degrees from bow) None
Receives Filtered Water False
 

Maintenance & Support

Sensor Log

    Date (UTC)     Submitter     Event Type     Details        
Nov 06 2024 00:26:47     Nahorniak, Jasmine     sensor specs update     logging_modes changed from [port,underway,transit,skeleton] to [file/db]    
Jan 02 2024 19:45:20     Clarke, Bonny     renamed sensor     Changed sensor ID from epp_rmr_000101 to epprmr_rad_000101.    
Dec 14 2023 18:36:03     Clarke, Bonny     renamed sensor     Changed sensor ID from radiom000101 to epp_rmr_000101.    
Dec 13 2023 17:47:04     Clarke, Bonny     sensor specs update     location_text changed from [Starboard Side] to [Crane Cab]    
Dec 13 2023 17:47:04     Clarke, Bonny     sensor specs update     enabled changed from [False] to [True]    
Dec 13 2023 17:47:04     Clarke, Bonny     sensor specs update     current_status changed from [calibration in progress] to [in use]    
Dec 13 2023 17:47:04     Clarke, Bonny     sensor specs update     current_location changed from [repair/calibration] to [installed]    
Dec 05 2023 18:56:59     Clarke, Bonny     sensor specs update     last_cal_date changed from [2022-11-15 00:00:00+00:00] to [2023-12-01 00:00:00+00:00]    
Dec 05 2023 18:56:59     Clarke, Bonny     sensor specs update     enabled changed from [True] to [False]    
Dec 05 2023 18:56:59     Clarke, Bonny     sensor specs update     current_status changed from [in use] to [calibration in progress]    
Dec 05 2023 18:56:59     Clarke, Bonny     sensor specs update     current_location changed from [installed] to [repair/calibration]    
Aug 01 2023 22:53:04     Nahorniak, Jasmine     sensor specs update     time_source changed from [vessel_time_server] to []    
Aug 01 2023 22:53:04     Nahorniak, Jasmine     sensor specs update     sensor_prefix changed from [] to [pyrgeo]    
May 17 2023 12:42:43     Clarke, Bonny     sensor specs update     last_cal_date changed from [None] to [2022-11-15 00:00:00+00:00]    
May 17 2023 12:42:43     Clarke, Bonny     sensor specs update     cal_recommended_freq_months changed from [None] to [12]    
Jun 10 2022 18:48:27     Romsos, Chris     sensor specs update     transmit_port changed from [16012] to [17012]    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     maintenance_C_text changed from [ ] to []    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     maintenance_B_text changed from [ ] to []    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     maintenance_A_text changed from [ ] to []    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     transmit_ip changed from [ ] to []    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     software_configuration changed from [ ] to []    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     software_version changed from [ ] to []    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     condition_details changed from [ ] to []    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     acquire_condition changed from [ ] to []    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     analog_units changed from [ ] to []    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     power_current_type changed from [] to [AC]    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     nmea2000 changed from [None] to [[]]    
May 19 2022 19:58:49     Romsos, Chris     sensor specs update     nmea0183 changed from [None] to [[]]    
May 19 2022 19:58:48     Romsos, Chris     sensor specs update     binary_format_string changed from [ ] to []    
May 19 2022 19:58:48     Romsos, Chris     sensor specs update     native_data_description changed from [ ] to []    
May 19 2022 19:58:48     Romsos, Chris     sensor specs update     sensor_ethernet_protocol changed from [ ] to []    
May 19 2022 19:58:48     Romsos, Chris     sensor specs update     sensor_ip changed from [ ] to []    
May 19 2022 19:58:48     Romsos, Chris     sensor specs update     comm_type changed from [ ] to []    
May 19 2022 19:58:48     Romsos, Chris     sensor specs update     signal_type changed from [ ] to []    
May 19 2022 19:58:48     Romsos, Chris     sensor specs update     model changed from [PIR-PSP] to [Eppley Precision Infrared Radiometer Pyrgeometer]    
May 19 2022 19:58:48     Romsos, Chris     sensor specs update     vendor changed from [The Eppley Laboratory, Inc.] to [The Eppley Laboratory, Inc.]    
Aug 07 2020 00:36:44     Nahorniak, Jasmine     sensor specs update     sensor_orientation changed from [] to [vertical up]    
 
Calibration & Maintenance

Frequency Last Completed Next Due
Maintenance A:
 
Maintenance B:
 
Maintenance C:
 
Calibration
  12 months Dec 01 2023 Dec 01 2024
Recommended Replacement
 
Warranty
 
 
Calibration Files

Date Version Author Type Document Access
Dec 01 2023           cal_file   PIR 35116F3 2023 cal
Dec 01 2023           cal_file   PSP 35149F3 2023cal
Nov 11 2022           cal_file   35116 2022 cal
Nov 11 2022           cal_file   35149 2022 cal
Dec 14 2021           cal_file   RAd 201 2021 cal
Dec 07 2021           cal_file   35149 2020 cal
Dec 09 2020           cal_file   35116 2020 cal
Jan 15 2018           cal_file   35116 2017 cal
Jan 12 2017           cal_file   35149 2017 cals
 
 
User Guides & Documents

Date Version Author Type Document Access
        The Eppley Laboratory, Inc.   brochure   EPLAB PIR Brochure
Jan 01 2001   2RA-A     Campbell Scientific, Inc.   application_note   Eppley PSP Application Note
Jun 01 1992   6/92     Campbell Scientific, Inc.   user_manual   Eppley PSP Instruction Manual
May 01 2006        The Eppley Laboratory, Inc.   instruction_sheet   Eppley PSP Instruction Sheet
        The Eppley Laboratory, Inc.   sensor_image   PIR Photo
 
 
Ownership

Quantity Value
Owner URI
Responsible Party URI Marine Technicians
Purchase Date
Acquire Date
Acquire Condition
End/Lost Ownership Date
 
Vendor Support

Support Type Details
Company The Eppley Laboratory, Inc.
Website http://www.eppleylab.com/
Phone +1 401 847 1020
Email info@eppleylab.com
Shipping The Eppley Laboratory, Inc., 12 Sheffield Avenue, Newport, RI 02840
 
Copyright Notice: All sensor-specific images and documentation on this page are copyright of the vendor.
 
 
OSU logo NSF logo RCRV logo
RCRV Datapresence and Engineering Support Center
Oregon State University
Corvallis, OR 97331, USA