Radiometer

Summary



TypeRadiometer
VendorThe Eppley Laboratory, Inc.
ModelEppley Precision Infrared Radiometer Pyrgeometer
S/NPIR30606F3; PSP30600F3
ID Prefixpyrgeo
CORIOLIX IDepprmr_rad_000102
ClassAtmospheric
LocationCrane Cab
Conditionunknown
Configurationdefault
Native Sample Rate0.25 Hz
UDP Port(s)None 16000
EnabledTrue
 
 
 

General Description



Two radiometers measuring irradiance include a pygeometer (PIR) measuring downwelling longwave irradiance and a pyranometer (PSP) measuring global shortwave irradiance.

 
 

Data Collection Specs

Native Data Feed
Signal Type
Communication Type
Serial Details
Baud Rate 19200
Parity N
Stop Bits 1
Byte Size 8
Timeout (s) None
Polled Start Commands
Polled Serial Commands
Ethernet Details
MAC Address
IP Address
Protocol
Source File Details
Path
Filename
Native Data Format
Native Data Description
Native Data Type text
Start of Message Characters
End of Message Characters
Sample Data Message $WIR01,19/01/17,19:18:32, 67, -285.2, 216.54, -0.44, -0.05, 405.76, 1.4, 10.9
Message Parser Format ['^\WWIR01,(?P<instrument_date>\d\d/\d\d/\d\d),(?P<instrument_time>\d\d:\d\d:\d\d),(?P<test>[\w])?\s*(?P<samples_in_average>\d+),\s+(?P<pir_volts>\-?\d+\.\d+),\s+(?P<pir>\-?\d+\.\d+),\s+(?P<pir_case_temp_false>\-?\d+\.\d+),\s+(?P<dome_temp>\-?\d+\.\d+),\s+(?P<psp>\-?\d+\.\d+),\s+(?P<diagnostic_rad_board_temp>\-?\d+\.\d+),\s+(?P<diagnostic_rad_board_vbat_voltage>\-?\d+\.\d+)$']
Data Processing
Data Processing Script NoTransformation
Data Rates
Native Sample Rate 0.25 Hz
Shipboard Network Packet Rate 0.25 Hz
Ship to Shore Rate 0.25 Hz
Data Storage & Access
Raw Data Ethernet Distribution
IP Address
Port None
Packet Size None
Transformed Data Ethernet Distribution
IP Address 192.167.0.255
Port 16000
Archive to Database True
Transmit Real-time Data to Shore True
Time Source
Source
 

Parameters


Quantity Infrared Irradiance (W/m2) Shortwave Irradiance (W/m2)
Plot Color
 
 
Parameter Full Name Infrared Irradiance Shortwave Irradiance
Parameter Short Name Infrared Irradiance Shortwave Irradiance
Parameter Standard Name XXXXXXXX XXXXXXXX
SAMOS Data Category Longwave Radiation Shortwave Radiation
Description Pyrgeometer measures the downwelling longwave (infrared) irradiance from the sky. The spectral range covered is 4 to 50 µm. The accuracy is 5 W/m2. Pyranometer (PSP) mounted in the horizontal plane to measure shortwave irradiance from 285 to 2800 nm.
Processing Status raw raw
Processing Symbol pir psp
Diagnostic False False
IOOS Category Meteorology Meteorology
Units W/m2 W/m2
Data Type double double
Gross Minimum Value None W/m2 None W/m2
Gross Maximum Value None W/m2 None W/m2
Recommended Minimum Value None W/m2 None W/m2
Recommended Maximum Value None W/m2 None W/m2
Global Minimum Value 0.0 W/m2 0.0 W/m2
Global Maximum Value 500.0 W/m2 1400.0 W/m2
Local Minimum Value None W/m2 None W/m2
Local Maximum Value None W/m2 None W/m2
Lower Detection Limit 2.0 W/m2 2.0 W/m2
Saturated Value 500.0 W/m2 1400.0 W/m2
Bad Data Value 500.0 W/m2 1400.0 W/m2
Resolution None W/m2 None W/m2
Accuracy +/- None W/m2 +/- None W/m2
Precision None W/m2 None W/m2
Binning Type linear linear
Online Display Format 5.1f 6.1f
Logging Modes
 

Physical Sensor Specs

Operating Limits
Input Power None - None V AC
Ideal Power None V AC
Operating Current None - None mA @ None V AC
Analog Output None - None
Operating Flow Rate None - None ml s-1
Ideal Flow Rate None ml s-1
Operating Depth None - None m
Operating Temperature None - None degree_C
Lamp Life None days
Response Time None seconds
Additional Time Offset 0.0 seconds
Physical Status
Current Location installed
Current Status in use
Physical Condition unknown
Physical Configuration default
Software Version
Software Configuration
Data Acquisition Device ID
Data Acquisition Device Serial Port
Data Acquisition Device Analog/Digital Port(s)
Data Acquisition Device Modbus Unit None
Sensor Use Level
Enabled True
Vessel Installation Location
Vessel Endeavor
Vessel Location Crane Cab
Location Coordinates (meters)
Sensor Orientation vertical up
Height Above Waterline (meters) None
Zero Line Reference (degrees from bow) None
Receives Filtered Water False
 

Maintenance & Support

Sensor Log

    Date (UTC)     Submitter     Event Type     Details        
Nov 06 2024 00:27:02     Nahorniak, Jasmine     sensor specs update     logging_modes changed from [port,underway,transit,skeleton] to [file/db]    
Jan 02 2024 19:45:00     Clarke, Bonny     renamed sensor     Changed sensor ID from epp_rmr_000102 to epprmr_rad_000102.    
Dec 14 2023 18:35:04     Clarke, Bonny     renamed sensor     Changed sensor ID from radiom000102 to epp_rmr_000102.    
Dec 13 2023 17:46:19     Clarke, Bonny     sensor specs update     location_text changed from [Starboard Side] to [Crane Cab]    
Dec 13 2023 17:46:19     Clarke, Bonny     sensor specs update     enabled changed from [False] to [True]    
Dec 13 2023 17:46:19     Clarke, Bonny     sensor specs update     current_status changed from [calibration in progress] to [in use]    
Dec 13 2023 17:46:19     Clarke, Bonny     sensor specs update     current_location changed from [repair/calibration] to [installed]    
Dec 05 2023 18:57:57     Clarke, Bonny     sensor specs update     last_cal_date changed from [2022-11-15 00:00:00+00:00] to [2023-12-01 00:00:00+00:00]    
Dec 05 2023 18:57:57     Clarke, Bonny     sensor specs update     current_status changed from [in use] to [calibration in progress]    
Dec 05 2023 18:57:57     Clarke, Bonny     sensor specs update     current_location changed from [installed] to [repair/calibration]    
Aug 01 2023 22:52:40     Nahorniak, Jasmine     sensor specs update     time_source changed from [vessel_time_server] to []    
Aug 01 2023 22:52:40     Nahorniak, Jasmine     sensor specs update     sensor_prefix changed from [] to [pyrgeo]    
Jun 20 2023 17:13:44     Nahorniak, Jasmine     sensor specs update     enabled changed from [True] to [False]    
May 17 2023 12:44:56     Clarke, Bonny     new sensor     Initial submission of sensor metadata.    
 
Calibration & Maintenance

Frequency Last Completed Next Due
Maintenance A:
 
Maintenance B:
 
Maintenance C:
 
Calibration
  12 months Dec 01 2023 Dec 01 2024
Recommended Replacement
 
Warranty
 
 
Calibration Files

Date Version Author Type Document Access
Dec 01 2023           cal_file   PIR 30606F3 2023 cal
Dec 01 2023           cal_file   PSP 30600F3 2023 cal
Nov 11 2022           cal_file   30600 2022 cal
Nov 11 2022           cal_file   30606 2022 cal
Dec 09 2021           cal_file   RAD 217 2021 cal
Dec 09 2020           cal_file   30606 2020 cal
Dec 07 2020           cal_file   30600 2020 cal
Jan 15 2018           cal_file   30606 2017 cal
Jan 12 2018           cal_file   30600 2017 cal
 
 
User Guides & Documents

Date Version Author Type Document Access
        The Eppley Laboratory, Inc.   brochure   EPLAB PIR Brochure
Jan 01 2001   2RA-A     Campbell Scientific, Inc.   application_note   Eppley PSP Application Note
Jun 01 1992   6/92     Campbell Scientific, Inc.   user_manual   Eppley PSP Instruction Manual
May 01 2006        The Eppley Laboratory, Inc.   instruction_sheet   Eppley PSP Instruction Sheet
        The Eppley Laboratory, Inc.   sensor_image   PIR Photo
 
 
Ownership

Quantity Value
Owner URI
Responsible Party URI Marine Technicians
Purchase Date
Acquire Date
Acquire Condition
End/Lost Ownership Date
 
Vendor Support

Support Type Details
Company The Eppley Laboratory, Inc.
Website http://www.eppleylab.com/
Phone +1 401 847 1020
Email info@eppleylab.com
Shipping The Eppley Laboratory, Inc., 12 Sheffield Avenue, Newport, RI 02840
 
Copyright Notice: All sensor-specific images and documentation on this page are copyright of the vendor.
 
 
OSU logo NSF logo RCRV logo
RCRV Datapresence and Engineering Support Center
Oregon State University
Corvallis, OR 97331, USA